Conguration
9.3 Scale
Up to 10 scales may be selected for the display of
the measurement results.
Select the desired scale to be changed and press
the operate button.
Following scales are available:
Longwave (0...100)
Shortwave (0...100)
The classical scales
Longwave (1,2 - 12 mm) and
Shortwave (0,3 - 1.2 mm)
DOI (Dorigon) (47...96)
Correlation to ASTM E 430, value range similar to
20° gloss.
DOI (GM) (47...96)
GMspecication
DOI (BYK) (30...100)
HigherresolutionbutlowervaluesthanDOI
(Dorigon)
du (1...60)
Dullness,nestructuressmallerthan0.1mm
Wa...Wd (0...100)
The structure spectrum is evaluated by applying
lterfunctionstothemeasuredopticalprole.
The values Wa...Wd stand for the respective
wavelength/range.