Advanced Operations
1. Press Nominal Values.
2. Press Edit.
An automatic menu guidance queries values (filters,
wavelength, nominal values and tolerances) given in the quality
control certificate, to the following specifications:
Stray Light
Photometrical accuracy
Wavelength accuracy
3. Press OK when all values are entered and the overview is
displayed.
4. Press Verification.
5. Insert the adapter Z (Figure 4 on page 11 ) in cell compartment
#2.
Vedere la pagina 73