C2 Colour Measurement
The colours of all transparent liquids can be measured, using 50mm square cuvettes or 11mm
round glass cuvettes, depending on their transmittances. If required, select another colour system
for printer output in the configuration menu (cp. page 19).
Instrument preparation
3. Turn instrument and printer on.
4. Calibrate the instrument as described in chapter. C1.
Measuring procedure
1. Fill the sample liquid into a cuvette and insert it into the corresponding compartment.
Measurement starts automatically.
2. After measuring, the colour values of the sample will
appear on the display
and be printed, e.g.
2.6 Iodine color value
507 Hazen color value
3. Now press the -key (DIM) until the desired colour
system appears on the display.
4. The colour system will be displayed during the
measurements to follow.
2.6
iodine color value
_DIM ¢Conf 11°
507
Hazen color value
_DIM ¢Conf 11°