Disconnect the instrument from
the compressed-air supply.
Remove the 4 cross-head rear
panel screws.
Slightly tilt the raer panel with
holder backwards and remove it
by carefully pulling it up and
guiding it beyound the
counterweight.
Remove the counterweight from
the panel support.
Loosen the 2 screws below the
panel support using a metric 2.5
mm hexagon socket screw key.
Carefully raise and remove the
panel support so that the spring
and the impact tool lie bare on the
guide tube.
Carefully remove the impact tool
and the spring.
Hold the instrument upside-down
in order to remove the ball from
the guide tube.
7. Replacing the
Impact Tool
Check the ball for any defects.
Then put the ball back in the
guide tube or replace it
respectively.
Put the spring in place.
Check the impact tool for any
defects and replace it if
necessary.
Put the impact tool in place.
Re-assemble all other parts in
reverse order.
Vedere la pagina 23