color-view™ Spectrophotometer User’s Guide
60
NOTE
It is important to remember not to touch the bulb or the inside
surface of the lamp reflector. Hand oils or dirt will cause
damage to these surfaces and render the lamp assembly useless.
8 Lift lamp out of its operation position and set on Sensor.
9 Locate end of lamp lead wires (yellow) which are
fastened to terminal block.
10 Loosen the retaining screws (which attach the wires to the
block) just enough so that the contact lugs can be easily
removed from under the screws. Do not completely
remove the screws from the block.
11 Remove the lamp from the compartment.
Replacement
1 Locate the spare replacement lamp in the accessory box.
2 Insert the contact lugs of the replacement lamp between
the screw head and the terminal block. Tighten the
screws to insure a good contact.
3 Insert the replacement lamp in the holder slot in the lamp
compartment and refasten wire spring.
4 Reinstall lamp compartment door.